The Pioneer PX Range is designed for simple, intelligent operation offers competitive performance and has been durably constructed for long-term use. This combination allows high accuracy and repeatability for applications in laboratory and education settings.
Each model is intuitively designed for intelligent operation and offers high accuracy and repeatability for essential weighing applications, as well as featuring a second line display for additional information or guidance, a static removal bar for convenient grounding, and USB and RS232 connectivity for easy communication. With cast metal lower housing, sub-pan and stainless steel weighing pan, the Pioneer Range is durably constructed for versatile, long-term use.
- Capacity: 3200g, 5200g and 4200g
- Readability: 10mg (3200g & 5200g) and 100mg (4200g)
- External and internal Calibration: InCal™ semi-automatic
- Applications: Basic Weighing, Parts Counting, Percent Weighing, Animal Weighing, Density Determination
- Display: Backlit Liquid Crystal Display (LCD); Second line display for additional information or guidance (dot matrix)
- Operation: AC adapter (included)
- Communication: RS232 and USB device (included); GLP/GMP data output with real-time clock
- Construction: Metal base, plastic top housing, removable stainless-steel pan, integrated weigh-below-hook, security bracket, calibration lock and in-use cover
- Design Features: User-selectable environmental filters and brightness settings, auto-tare, auto-dim, user-selectable span calibration points, software lockout and reset menu, user-selectable communication settings and data print options, user-definable project and user IDs, software overload/underload indicator, stability indicator, four operating languages
InCal™ Model | PX3202 | PX5202 | PX4201 |
ExCal Model | PX3202/E | PX5202/E | PX4201/E |
Approved Model* | PX3202M | PX5202/M | PX4201/M |
Verification Interval (e) (g) | 100mg | 100mg | 1g |
Capacity (g) | 3200 | 5200 | 4200 |
Readability (g) | 10mg | 10mg | 100mg |
Repeatability (STDEV) (g) | 10mg | 10mg | 100mg |
Linearity (g) | ±20mg | ±0.2mg | ±0.2mg |
Stabilization Time (s) | 1 | 1 | 1 |
Sensitivity Temperature Drift (PPM/K) | ±3 | ±3 | ±10 |
Typical Minimum Weight USP (USP K=2, U=0.10%) | 20g | 20g | 200g |
Pan Size (Ø) | 180mm | 180mm | 180mm |
Units | Milligram, Gram, Kilogram, Ounce, Pound, Carat, Pennyweight, Ounce Troy, Grain, Newton, Hong Kong Tael, Singapore Tael, Taiwan Tael, Momme, Tical (MM), Mesghal, Tola (India), 1 custom unit |
Units (Approved Models) | g, mg , ct |
Applications | Basic Weighing, Parts Counting, Percent Weighing, Animal Weighing, Density Determination |
Power Supply | Power Input: 100 - 240V ~ 200mA 50 - 60Hz 12 - 18VA Power Output: 12 VDC 0.5A |
Communication | RS232 and USB |
Operating Temperature Range | 10°C to 30°C |
Net Weight | 3.5kg |
Balance Dimensions | 209 × 321 × 98 mm |
Shipping Weight | 5kg |
Shipping Dimensions (W × D × H) | 550 × 385 × 291 mm |
Compliance
Metrology (PX models only): EN 45501; OIML R76
Product Safety: IEC/EN 61010-1; CAN/CSA C22.2 61010-1; UL 61010-1
Compliance Marks: CE; CSA